6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
:CALCulate:LLINe[1]|2|...6:DISPlay?
Example
:CALC:LLIN2:DISP ON turns on the display for limit line 2.
Dependencies
This command will generate an “Option not available” error message unless you have the proper
option installed in your instrument.
Couplings
Limit display ON selects the limit.
Testing is done on all displayed limits if Test Limits (All Limits) is ON.
Entering the limit menu from the GUI turns on the selected limit.
Preset
OFF
State Saved
Saved in instrument state.
Backwards
Compatibility SCPI
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1
In the past you had to send the DISP command as well as the STATe command in order to get a
limit on and testing. Now, the DISP command is sufficient, but we accept the state command
and map it to DISP
Initial S/W Revision
A.02.00
Properties
Accesses a menu which lets you set the properties of the selected limit.
Key Path
Meas Setup, Limits
Initial S/W Revision
A.02.00
Select Limit
Specifies the selected limit. The term “selected limit” is used throughout this document to specify
which limit will be affected by the functions.
Key Path
Meas Setup, Limits, Properties
Notes
The selected limit is remembered even when not in the Limit Menu.
Preset
Limit 1, not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00