5 Legacy Command Descriptions
CLRW (Clear Write)
The following table describes the parameters set by N9061A in the X-series instrument when the
legacy commands for trace mode and averaging are sent.
Legacy Products
X-series
Trace Commands
Averaging (VAVG)
Detector
(DET)
Trace/Detector
Trace Type
View/Blank
Detector
CLRW
Off
Normal
ClearWrite
On
Last set
CLRW
On
Sample
Trace Average
On
Sample
MXMH
Off
Peak
Max Hold
On
Peak
MXMH
On
Sample
Trace Average
On
Peak
MINH
Off
NegPeak
Min Hold
On
NegPeak
MINH
On
Sample
Trace Average
On
NegPeak
VIEW
Off
Normal
No change
View
No change
VIEW
On
Sample
Trace Average
View
Sample
BLANK
Off
Normal
No change
Blank
No change
BLANK
On
Sample
Trace Average
Blank
Sample
Legacy Products command mapping to X-series for trace/detector settings
For example, if an 8560 series analyzer receives CLRW, and averaging is set to ON, then the
analyzer's detector is automatically set to Sample. In the same circumstances, N9061A sets the X-
series instrument trace type to Trace Average, View/Blank to On, and the Detector to Sample.
Format
CLRW TRA|TRB
Preset: CLRW TRA
Query Data Type
N/A
SCPI Equivalent Commands
:TRACe1|2|3|4|5|6:TYPE WRITe (see
Preset
TRA (after a preset, only trace A is set to clearwrite)
Notes
The functions of CLRW are identical to the command
"A1 [one] (Clear Write for Trace A)"
and
"B1 [one] (Clear Write for Trace B)" on page 242
.
264
Remote Language Compatibility Measurement Application Reference