Mode setup, Hp8560 series, hp8566/68 – Agilent Technologies Signal Analyzer N9030a User Manual
Page 886
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Mode Setup
Mode Setup
This menu allows you to select 8560 series analyzer emulation, or 8566A/B, 8568A/B emulation,
and hence the remote control command set that will be recognized by the X-Series instrument.
Key Path
Front-panel Key
Remote Command
:SYSTem:LANGuage HP8560E | HP8561E | HP8562E |
HP8563E | HP8564E | HP8565E | HP8566A | HP8566B |
HP8568A | HP8568B
:SYSTem:LANGuage?
Default
HP8563E
HP8560 series, HP8566/68
These keys have submenus that allow you to select which legacy analyzer to emulate. The selected
analyzer type determines the response to the ID? command, and affects the behavior of
commands such as IP. You can use any command offered by any of the legacy analyzers,
regardless of the language setting. However, if the command is not correct for the selected legacy
analyzer, the command may not work as expected.
The legacy analyzer emulation selection in this menu does not affect the response to the SCPI
query *IDN?. This query still returns the model number and firmware version number of the X-
Series instrument.
The legacy instrument selections are as follows:
Key
Response to Command ID?
Notes
8560E/EC
HP8560E
Selects 8560E/EC remote programming language
8561E/EC
HP8561E
Selects 8561E/EC remote programming language
8562E/EC
HP8562E
Selects 8562E/EC remote programming language
8563E/EC
HP8563E
Selects 8563E/EC remote programming language
This is the default selection for the N9061A application.
8564E/EC
HP8564E
Selects 8564E/EC remote programming language
8565E/EC
HP8565E
Selects 8565E/EC remote programming language
HP8566A
HP8566A
Selects HP8566A remote programming language
HP8566B
HP8566B
Selects HP8566B remote programming language
HP8568A
HP8568A
Selects HP8568A remote programming language
HP8568B
HP8568B
Selects HP8568B remote programming language
886