6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Key Path
Trigger, Auto/Holdoff
Remote Command
:TRIGger[:SEQuence]:HOLDoff
:TRIGger[:SEQuence]:HOLDoff?
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:HOLDoff:STATe?
Example
TRIG:HOLD:STAT ON
TRIG:HOLD 100 ms
Dependencies
Unavailable if the selected Input is BBIQ. If this is the case, the key is grayed out if it is pressed
the informational message “Feature not supported for this Input” is displayed. If the SCPI
command is sent, the error “Settings conflict; Feature not supported for this Input” is generated.
Preset
Off, 100 ms
State Saved
Saved in instrument state
Min
0 s
Max
0.5 s
Default Unit
s
Initial S/W Revision
Prior to A.02.00
1438
Remote Language Compatibility Measurement Application Reference