6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
No confirmation is required when Alert is configured through a remote command.
Key Path
System, Alignments, Auto Align, Alert
Mode
All
Example
:CAL:AUTO:ALER NONE
Initial S/W Revision
Prior to A.02.00
Execute Expired Alignments (Remote Command Only)
Alignments can be expired in the situation where Auto Align is in the state of Partial or Off. This
feature runs the alignments that have expired. This is different than performing an Align All, Now
operation, which performs an alignment of all subsystems, regardless of whether they are needed
or not. Execute Expired Alignments performs alignments of only the individual subsystems that
have become due.
Mode
All
Remote Command
:CALibration:EXPired?
Example
:CAL:EXP?
Notes
:CALibration:EXPired? returns 0 if successful
:CALibration:EXPired? returns 1 if failed
Initial S/W Revision
Prior to A.02.00
Align Now
Accesses alignment processes that are immediate action operations. They perform complete
operations and run until they are complete.