Bxc (exchange trace b and trace c), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 251: Description, Syntax legacy products
5 Legacy Command Descriptions
BXC (Exchange Trace B and Trace C)
BXC (Exchange Trace B and Trace C)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Exchanges Trace B data with Trace C data.
Trace C must not be an active trace. This means that the data in Trace C cannot be updated as
the instrument sweeps. To ensure that the current settings of the instrument are reflected in the
data exchanged between Trace B and Trace C, you must follow the four step process below.
1. Select single sweep mode (S2 or SNGLS command)
2. Select the desired instrument settings
3. Take one complete sweep
4. Exchange the data
Format
BXC
Query Data Type
N/A
SCPI Equivalent Commands
TRACe3:TYPe?
TRACe3:UPDate?
TRACe3:DISPlay?
TRACe2:TYPe?
TRACe2:UPDate?
TRACe2:DISPlay?
TRACe:EXCHange TRACE2, TRACE3
(See
, etc.)
Couplings
Trace Update is set to Off and Trace Display is set to On.
Notes
The functions of BXC are identical to the command
"KSi (Exchange Trace B and Trace C)
and to the XCH TRB,TRC form of the command
.
Remote Language Compatibility Measurement Application Reference
251