5  Legacy Command Descriptions

BXC (Exchange Trace B and Trace C)

BXC (Exchange Trace B and Trace C)

Syntax

Legacy Products

8566A/B, 8568A/B

Description

Exchanges Trace B data with Trace C data.

Trace C must not be an active trace. This means that the data in Trace C cannot be updated as
the instrument sweeps. To ensure that the current settings of the instrument are reflected in the
data exchanged between Trace B and Trace C, you must follow the four step process below.

1. Select single sweep mode (S2 or SNGLS command)

2. Select the desired instrument settings

3. Take one complete sweep

4. Exchange the data

Format

BXC

Query Data Type

N/A

SCPI Equivalent Commands

TRACe3:TYPe?

TRACe3:UPDate?

TRACe3:DISPlay?

TRACe2:TYPe?

TRACe2:UPDate?

TRACe2:DISPlay?

TRACe:EXCHange TRACE2, TRACE3

(See

"View/Blank " on page 1344

, etc.)

Couplings

Trace Update is set to Off and Trace Display is set to On.

Notes

The functions of BXC are identical to the command

"KSi (Exchange Trace B and Trace C)

" on page 334

and to the XCH TRB,TRC form of the command

"XCH (Exchange)" on

page 509

.