6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
Preset
Marker 1
State Saved
The number of the selected marker is saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Select Marker
Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.
Key Path
Marker
Notes
The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.
Preset
Marker 1
State Saved
The number of the selected marker is saved in instrument state.
Initial S/W Revision
Prior to A.02.00
X Axis Scale
Accesses a menu that enables you to affect how the X Axis information for the selected marker is
displayed in the marker area (top-right of display) and the active function area of the display,
and how the marker is controlled. The available settings for the X Axis Scale are Frequency,
Period, Time, and Inverse Time.
See
"More Information" on page 695
.
Key Path
Marker, Properties
Remote Command
:CALCulate:MARKer[1]|2|...12:X:READout FREQuency | TIME | ITIMe |
PERiod
:CALCulate:MARKer[1]|2|...12:X:READout?
:CALCulate:MARKer[1]|2|...12:X:READout:AUTO ON|OFF|1|0
:CALCulate:MARKer[1]|2|...12:X:READout:AUTO?
Example
CALC:MARK3:X:READ TIME sets the marker 3 X Axis Scale to Time.
Notes
This command causes the specified marker to become selected.
Preset
AUTO
Marker Preset (selected when a marker is turned Off): Auto (see below). In most measurements
the Auto settings results in Frequency being the preset readout.