6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
This key is grayed out in modes that do not support BTS.
Preset
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"
State Saved
Saved in instrument state.
Min
–100 dB
Max
100 dB
Readback
BTS,
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:CORRection:BTS[:RF]:LOSS
[:SENSe]:CORRection:BTS[:RF]:LOSS?
Example
CORR:BTS:LOSS 10 sets the Ext Gain value to –10 dB, and subsequently querying :LOSS will
give 10 dB
CORR:BTS:LOSS –10 sets the Ext Gain value to 10 dB, and subsequently querying :LOSS will give
–10 dB
Notes
A positive value of
indicates a gain.
Anytime :LOSS is set it sets :GAIN to the negative value of the parameter sent.
Anytime :LOSS is queried it gives the negative of :GAIN
Preset
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"
Min
100 dB
Max
–100 dB
Initial S/W Revision
Prior to A.02.00
I Ext Gain
This function affects only the I channel input, except when the Input Path is I+jQ. In I+jQ this
setting is applied to both I and Q channel inputs. It is not available unless the Baseband I/Q
option (BBA) is installed.
Key Path
Input/Output, External Gain
Remote Command
[:SENSe]:CORRection:IQ:I:GAIN
[:SENSe]:CORRection:IQ:I:GAIN?
622