6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Input/Output

Output Config

Accesses keys that configure various output settings, like the frequency reference output, trigger
output and analog output.

Key Path

Input/Output

Backwards Compatibility
Notes

In ESA there was not a user interface to enable the Video Output (Analog Output), Trigger

Output, or Gate Output. In the X-Series each of these physical connectors requires
configuration, thus the user interface has been added for X-Series, along with the potential for
an output you think is always on to be switched off.

Initial S/W Revision

Prior to A.02.00

Trig Out (1 and 2)

Select the type of output signal that will be output from the rear-panel Trig 1 Out or Trig 2 Out
connectors.

Key Path

Input/Output, Output Config

Remote Command

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut HSWP | MEASuring | MAIN |
GATE | GTRigger | OEVen | SPOint | SSWeep | SSETtled | S1Marker |
S2Marker | S3Marker | S4Marker | OFF

:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut?

Example

TRIG:OUTP HSWP

TRIG2:OUTP GATE

Dependencies

The second Trigger output (Trig 2 Out) does not appear in all models; in models that do not
support it, the Trig 2 Out key is blanked, and sending the SCPI command for this output
generates an error, “Hardware missing; Not available for this model number”   In models that do
not support the Trigger 2 output, this error is returned if trying to set Trig 2 Out and a query of
Trig 2 Out returns OFF.

Preset

Trigger 1: Sweeping (HSWP)

Trigger 2: Gate

This is unaffected by a Preset but is preset to the above values on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"

State Saved

Saved in instrument state

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference