Output config – Agilent Technologies Signal Analyzer N9030a User Manual
Page 649
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Output Config
Accesses keys that configure various output settings, like the frequency reference output, trigger
output and analog output.
Key Path
Input/Output
Backwards Compatibility
Notes
In ESA there was not a user interface to enable the Video Output (Analog Output), Trigger
Output, or Gate Output. In the X-Series each of these physical connectors requires
configuration, thus the user interface has been added for X-Series, along with the potential for
an output you think is always on to be switched off.
Initial S/W Revision
Prior to A.02.00
Trig Out (1 and 2)
Select the type of output signal that will be output from the rear-panel Trig 1 Out or Trig 2 Out
connectors.
Key Path
Input/Output, Output Config
Remote Command
:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut HSWP | MEASuring | MAIN |
GATE | GTRigger | OEVen | SPOint | SSWeep | SSETtled | S1Marker |
S2Marker | S3Marker | S4Marker | OFF
:TRIGger|TRIGger1|TRIGger2[:SEQuence]:OUTPut?
Example
TRIG:OUTP HSWP
TRIG2:OUTP GATE
Dependencies
The second Trigger output (Trig 2 Out) does not appear in all models; in models that do not
support it, the Trig 2 Out key is blanked, and sending the SCPI command for this output
generates an error, “Hardware missing; Not available for this model number” In models that do
not support the Trigger 2 output, this error is returned if trying to set Trig 2 Out and a query of
Trig 2 Out returns OFF.
Preset
Trigger 1: Sweeping (HSWP)
Trigger 2: Gate
This is unaffected by a Preset but is preset to the above values on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
649