Agilent Technologies Signal Analyzer N9030a User Manual
Page 1344
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
•
if the trace is the target of a Copy or participant in an Exchange
•
if the trace is cleared using the Clear Trace function (below)
When a trace becomes active (Update=On), the trace is cleared, the average count is reset, and a
new sweep is initiated.
Traces which are blanked (Display=off) do not display nor appear on printouts but are otherwise
unaffected. They may be queried and markers may be placed on them.
Note that the action of putting a trace in Display=Off and/or Update=Off does not restart the
sweep and does not restart Averaging or Hold functions for any traces.
Note also that whenever you set Update to On for any trace,Display is set to On for that trace.
View/Blank
This key lets you set the state of the two trace variables, Update and Display. The four choices
available in this 1-of-N menu are:
•
Trace On: Update and Display both On
•
View: Update Off and Display On
•
Blank: Update Off and Display Off
•
Background: Update On, Display Off (this allows a trace to be blanked and continue to update
“in the background”, which was not possible in the past)
A trace with Display Off is indicated by a strikethrough thru the type letter in the trace
annotation panel in the Measurement bar. A trace with Update Off is indicated by dimming the
type letter in the trace annotation panel in the Measurement bar. So in the example below, Traces
3, 4, 5 and 6 have Update Off and Traces 4 and 6 have Display Off.
See
"Trace Update State On/Off" on page 1345
See
"Trace Display State On/Off" on page 1346
.
See
"More Information" on page 1346
Key Path
Trace/Detector
Notes
The four states of this 1-of-N actually set two variables, Update and Display, to their four
1344
Remote Language Compatibility Measurement Application Reference