6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall
“(empty)" if no prior save operation has been performed to this register.
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
Prior to A.11.00
Register 1 thru Register 16
Selecting any one of these register keys causes the State of the mode from the specified Register to
be recalled. Each of the register keys annotates whether it is empty or at what date and time it
was last modified. In addition, you can use the Edit Register Names key under Save, State to enter
custom names for each register.
Registers are shared by all modes, so recalling from any one of the registers will cause a mode
switch to the mode that was active when the save to the Register occurred.
Although these 16 registers are the only registers available from the front panel, there are 128
state registers available in the instrument. Registers 17–128 are only available from the SCPI
interface, using the *RCL command.
After the recall completes, the message "Register
message bar. If you are in the Spectrum Analyzer Mode, and you are recalling a register that was
saved in the Spectrum Analyzer Mode, then after the recall, you will still be in the Recall Register
menu. If the Recall causes you to switch modes, then after the Recall, you will be in the
Frequency menu.
If a requested register is empty an error is generated.
Key Path
Recall, State
Example
*RCL 1
Range
1–16 from front panel, 1–128 from SCPI
Readback
Date and time with seconds resolution are displayed on the key
OR
A custom name of up to 30 characters entered using the Save, State,Edit Register Names key
OR
“(empty)" if no prior save operation has been performed to this register.
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
Prior to A.11.00
960
Remote Language Compatibility Measurement Application Reference