6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Delete All Corrections
Erases all correction values for all 4 Amplitude Correction sets.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK
key to delete all corrections. Press ESC or Cancel to close this dialog.” The deletion is only
performed if you press OK or Enter.
Key Path
Input/Output, Corrections
Remote Command
[:SENSe]:CORRection:CSET:ALL:DELete
Example
CORR:CSET:ALL:DEL
Initial S/W Revision
A.02.00
Merge Correction Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value
separated by commas. The difference between this command and Set Data is that this merges new
correction points into an existing set.
Any new point with the same frequency as an existing correction point will replace the existing
point’s amplitude with that of the new point.
An Ampcor array can contain a maximum of 2000 points.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...6:DATA:MERGe
Example
CORR:CSET1:DATA:MERGE 15000000, –5.0, 25000000, 5.0
This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to whatever values
already exist in correction set 1.
Preset
Empty after Restore Input/Output Defaults. Survives shutdown/restart of analyzer application
(including power cycle)
Min
Freq: 0 Hz
Amptd: –1000 dBm
Max
Freq: 1 THz
Amptd: +1000 dBm
Initial S/W Revision
A.02.00
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value
separated by commas.