beautypg.com

Swpcpl (sweep couple), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 476: Description, Syntax legacy products

background image

5  Legacy Command Descriptions
SWPCPL (Sweep Couple)

SWPCPL (Sweep Couple)

Syntax

Legacy Products

8560 series

Description

Selects either a stimulus-response (SR) or signal-analyzer (SA) auto-coupled sweep time. In
stimulus response mode, auto-coupled sweep times are usually much faster for swept response
measurements. Stimulus response auto-coupled sweep times are typically valid in stimulus-
response measurements when the system frequency span is less than 20 times the bandwidth of
the device under test.

Format

SWPCPL SA|SR

SWPCPL?

Query Data Type

SA|SR

SCPI Equivalent Commands

[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy|SRESponse

[:SENSe]:SWEep:TIME:AUTO:RULes?

(See

"Sweep Time Rules" on page 1154

)

Preset

SA

476

Remote Language Compatibility Measurement Application Reference