DET (Detection Mode)
DET (Detection Mode)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Selects the type of instrument detection (NEGative peak, NoRMal, POSitive peak, or SaMPle).
NEG
Enables negative peak detection.
NRM
Enables the ‘rosenfell’ detection algorithm that selectively chooses between positive and negative
values.
POS
Enables positive-peak detection, which displays the maximum video signal detected over a number of
instantaneous samples for a particular frequency.
SMP
Enables sample detection, which uses the instantaneous video signal value. Video averaging and noise-
level markers, when activated, activate sample detection automatically.
Format
DET NEG|NRM|POS|SMP (For option descriptions, see table above)
DET?
Query Data Type
NEG|NRM|POS|SMP
SCPI Equivalent Commands
[:SENSe]:DETector[:FUNCtion] (NEGative|NORMal|POSitive|SAMPle)
[:SENSe]:DETector[:FUNCtion]?
(See
)
Preset
NRM
Notes
The functions of DET are identical to the commands
"KSa (Normal Detection)" on page
"KSb (Positive Peak Detection)" on page 323
,
"KSd (Negative Peak Detection)" on
, and
"KSe (Sample Detection)" on page 329
.
Remote Language Compatibility Measurement Application Reference