Agilent Technologies Signal Analyzer N9030a User Manual
Page 777
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
Center Presel On/Off
This key controls the automatic centering of the preselector for the Measure at Marker function.
When Center Presel is On, the first step in performing the Measure at Marker function is to
perform a Presel Center. This is not performed if the microwave preselector is off, or the selected
marker’s frequency is below Band 1. If the function is not performed, no message is generated.
Key Path
Marker Function, Measure at Marker
Remote Command
:CALCulate:MAMarker:PCENter ON|OFF|1|0
:CALCulate:MAMarker:PCENter?
Example
:CALC:MAM:PCEN ON
Dependencies
Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in these
instruments, it is accepted without error, and the query always returns 0.
Preset
On
Backwards
Compatibility SCPI
[:SENSe]:EMI:MEASure:PCENter[:STATe] OFF|ON|0|1
[:SENSe]:EMI:MEASure:PCENter[:STATe]?
This command is included for compatibility with the E7400 and PSA option 239 .
Initial S/W Revision
A.02.00
Remote Language Compatibility Measurement Application Reference