6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
FM
Pressing this key, when it is not selected, selects and activates the FM demodulation function.
Pressing it a second time branches to the FM Demod menu where FM demodulation settings can
be adjusted.
Key Path
Meas Setup, Analog Demod Tune&Listen
Example
DEM FM
State Saved
Saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Channel BW (FM Demod)
Sets the RBW setting used by the hardware during the demodulation period in nonzero spans.
Note that this is a separate parameter only for the demodulation function and does not affect the
RBW setting in the BW menu which is used during the normal sweep. The flat top filter type must
be used during the demodulation period. A 5 kHz Video Bandwidth filter is used.
In Zero Span, the instrument’s RBW & VBW filters are used for the demodulation; thus, the
Channel BW (and RBW filter type) will match those of the instrument. This allows gap-free
listening. The Channel BW key is grayed out and the value displayed on the key matches the
current RBW of the instrument. Upon leaving zero span, the previous setting of Channel BW and
the flattop filter type are restored.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM
Remote Command
[:SENSe]:DEMod:FM:BANDwidth:CHANnel
[:SENSe]:DEMod:FM:BANDwidth:CHANnel?
Example
DEM:FM:BAND:CHAN 200 MHz
Notes
This key / command is grayed out in zero span
Dependencies
Unavailable in zero span.
Couplings
In zero span only, the value is set equal to the instrument’s current RBW value and it displays
that value on the softkey, but the softkey is grayed out.
Preset
150 kHz
State Saved
Saved in instrument state.
Min
390 Hz
Max
8 MHz
Default Unit
Hz
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference