Agilent Technologies Signal Analyzer N9030a User Manual
Page 1278

6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TIME:LPReselector?
Example
:CAL:TIME:LPR?
Notes
Value is the date and time the last successful Characterize Preselector was executed. The date is
separated from the time by a space character. Returns “” if no Characterize Preselector has ever
been performed on the instrument.
Dependencies
In models that do not include preselectors, this command is not enabled and any attempt to set
or query will yield an error.
State Saved
No
Initial S/W Revision
Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TEMPerature:LPReselector?
Example
:CAL:TEMP:LPR?
Notes
Value is in degrees Centigrade at which the last successful Characterize Preselector was
executed.
Dependencies
In models that do not include preselectors, this command is not enabled and any attempt to set
or query will yield an error.
State Saved
No
Initial S/W Revision
Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:AUTO:TIME:OFF?
Example
:CAL:AUTO:TIME:OFF?
Notes
Value is the elapsed time, in seconds, since Auto Align has been set to Off or Off with Alert. The
value is 0 if Auto Align is ALL or NORF.
State Saved
No
Initial S/W Revision
Prior to A.02.00
1278