6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
3. If the new absolute RF Burst level differs from the previous by more than 0.5 dB, the new level
is sent to the hardware; otherwise it is not updated (to avoid slowing down the acquisition)
Steps 2 and 3 repeat for subsequent measurements.
Key Path
Trigger, RF Burst
Scope
Meas Global
Remote Command
:TRIGger[:SEQuence]:RFBurst:LEVel:RELative
:TRIGger[:SEQuence]:RFBurst:LEVel:RELative?
Example
TRIG:RFB:LEV:REL –10 dB
sets the trigger level of the RF burst envelope signal to the relative level of –10 dB
Notes
Sending this command does not switch the setting from absolute to relative; to switch it you
need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.
The relative trigger level is not available in some measurements. In those measurements the
RELative parameter, and the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
command (above), will generate an error if sent.
Dependencies
This key is grayed out and Absolute Trigger Level selected if the required hardware is not present
in your analyzer and the current measurement does not support Relative triggering.
Preset
–6 dB
GSM: –25 dB
State Saved
Saved in instrument state
Min
–45 dB
Max
0 dB
Default Unit
dB or dBc
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:RFBurst:LEVel
This legacy command is aliased to :TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the
PSA had ONLY relative burst triggering
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.04.00
Trigger Slope
It is set positive to trigger on a rising edge and negative to trigger on a falling edge.
Key Path
Trigger, RF Burst
Remote Language Compatibility Measurement Application Reference