Agilent Technologies Signal Analyzer N9030a User Manual
Page 171
3 Programming the Analyzer
STATus Subsystem
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Temperature Register
The following commands and queries are available for this register:
"Questionable Temperature Condition" on page 171
"Questionable Temperature Enable" on page 171
"Questionable Temperature Event Query" on page 172
"Questionable Temperature Negative Transition" on page 172
"Questionable Temperature Positive Transition" on page 173
Questionable Temperature Condition
This query returns the decimal value of the sum of the bits in the Questionable Temperature
Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:CONDition?
Example
STAT:QUES:TEMP:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Enable
This command determines which bits in the Questionable Temperature Condition Register will set
bits in the Questionable Temperature Event register, which also sets the Temperature Summary bit
(bit 4) in the Questionable Register.
The variable
Remote Language Compatibility Measurement Application Reference
171