Agilent Technologies Signal Analyzer N9030a User Manual
Page 1155
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Swept Sweep Type (either manually or automatically chosen) and not when in FFT sweeps.
See
"More Information" on page 1155
.
Key Path
Sweep/Control, Sweep Setup
Remote Command
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy|SRESponse
[:SENSe]:SWEep:TIME:AUTO:RULes?
Example
SWE:TIME:AUTO:RUL ACC
Dependencies
In Zero Span, this key is irrelevant and cannot be accessed (because the whole Sweep Setup
menu is grayed out in Zero Span), however its settings can be changed remotely with no error
indication.
Grayed out in FFT sweeps. Pressing the key while the instrument is in FFT sweep generates an
advisory message. The SCPI is acted upon if sent, but has no effect other than to change the
readout on the key, as long as the analyzer is in an FFT sweep.
Couplings
Set to Auto on Auto Couple
Preset
AUTO
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:SWEep:TIME:AUTO:MODE SRESponse
This legacy command is aliased to :SWEep:TIME:AUTO:RULes SRESponse
Backwards
Compatibility SCPI
:SWEep:TIME:AUTO:MODE SANalyzer
This legacy command is aliased to :SWEep:TIME:AUTO:RULes NORMal
Backwards
Compatibility SCPI
:SWEep:TIME:AUTO:MODE?
This legacy query is aliased to :SWEep:TIME:RULes?, so it will match for SRESponse but not for
SANalyzer
Backwards Compatibility
Notes
The old Auto Sweep Time command was the same
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy
so it still works although it now has a third parameter (SRESponse).
The old Sweep Coupling command was
[:SENSe]:SWEep:TIME:AUTO:MODE SRESponse|SANalyzer and it is aliased as below:
Initial S/W Revision
Prior to A.02.00
More Information
The first set of rules is called SA – Normal. Sweep Time Rules is set to SA-Normal on a Preset or Auto
Couple
. These rules give optimal sweep times at a lossof accuracy. Note that this means that in the
Preset or Auto Coupled state, instrument amplitude accuracy specifications do not apply.
Remote Language Compatibility Measurement Application Reference
1155