Agilent Technologies Signal Analyzer N9030a User Manual
Page 532
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
Key Path
AMPTD Y Scale
Scope
Meas Global
Remote Command
[:SENSe]:POWer[:RF]:PADJust
[:SENSe]:POWer[:RF]:PADJust?
Example
POW:PADJ 100KHz
POW:PADJ?
Notes
The value on the key reads out to 0.1 MHz resolution.
Dependencies
•
Grayed out if microwave preselector is off. )
•
Grayed out if entirely in Band 0.
•
Blank in models that do not include a preselector, such as option 503. If the SCPI is sent in
these instruments, it is accepted without error, and the query always returns 0.
•
Grayed out in the Spectrogram View.
Preset
0 MHz
State Saved
The Presel Adjust value set by Presel Center, or by manually adjusting Presel Adjust, is not
saved in instrument state, and does not survive a Preset or power cycle.
Min
–500 MHz
Max
500 MHz
Default Unit
Hz
Backwards
Compatibility SCPI
[:SENSe]:POWer[:RF]:MW:PADJust
[:SENSe]:POWer[:RF]:MMW:PADJust
PSA had multiple preselectors, but the X-Series has only one. These commands simply alias to
[:SENSe]:POWer[:RF]:PADJust
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
Remote Command
[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe | MMWave | EXTernal
[:SENSe]:POWer[:RF]:PADJust:PRESelector?
Notes
PSA had multiple preselectors, and you could select which preselector to adjust. Since the X-
Series has only one mm/uW preselector, the preselector selection softkey is no longer available.
However, to provide backward compatibility, we accept the legacy remote commands.
The command form has no effect, the query always returns MWAVe
Initial S/W Revision
Prior to A.02.00
532
Remote Language Compatibility Measurement Application Reference