6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
•
Set the Detector to AVERAGE
•
Set the sweep time to 16ms - sets the variance correctly for good results.
•
Set a Band/Interval Power Marker function and set the interval over the full width of trace i.e.
Left to 0s and Right to 16ms
Noise Source
This menu allows you to turn the noise source power on or off when making manual noise figure
measurements.
See
"More Information" on page 856
.
Key Path
Meas Setup
Remote Command
:SOURce:NOISe:TYPE NORMal | SNS
:SOURce:NOISe:TYPE?
Example
SOUR:NOIS:TYPE NORM
Couplings
If no SNS is connected, this parameter will be set to “Normal”
When Type is set to “SNS” and the SNS is disconnected, this parameter gets bumped to
“Normal”
When an SNS is not connected, the SNS type will be grayed (disabled).
Preset
Normal
State Saved
Saved in instrument state.
Range
Normal | SNS
Backwards Compatibility
Notes
In previous Noise Figure analysis applications, this command
could optionally be preceded with the :SENSe keyword. The
optional :SENSe keyword is no longer supported.
Initial S/W Revision
Prior to A.02.00
More Information
There are 2 types of noise sources: a Smart Noise Source (SNS), and a "Normal" noise source - e.g.
346 series. This menu allows the user to control both. The SNS has its own connector on the rear
of the analyzer and when it is connected the user can then select it from the “Type” 1 of N,
allowing the State parameter to then control the SNS. The "Normal" source is controlled by a BNC
connector that supplies 28V. If SNS is NOT connected then the “state” parameter controls the
"Normal" noise source 28V BNC port. If both are connected the “Type” parameter will determine
which source the “State” parameter will control. Two sources can never be controlled together. The
“SNS attached” SCPI query detailed below can be used remotely to determine if an SNS is
connected. SNS functionality is limited to turning on and off only. The SNS ENR data and
temperature cannot be queried, unless the Noise Figure application is installed. The SNS ENR