Agilent Technologies Signal Analyzer N9030a User Manual
Page 904
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Peak Search
The Peak Table is updated after each sweep. The list of peaks in the Peak Table can be ordered
either by ascending frequency or by descending amplitude. In either case, the entire trace is first
evaluated and the 20 highest peaks are selected for inclusion in the list. After the peaks are
selected, they are then sorted and displayed according to the Peak Sort setting.
Key Path
Peak Search
Initial S/W Revision
Prior to A.02.00
Peak Table On/Off
Turns Peak Table on/off. When turned on, the display is split into a measurement window and a
peak table display window.
Turning the Peak Table on turns the Marker Table off and vice versa.
Key Path
Peak Search, Peak Table
Remote Command
:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1
:CALCulate:MARKer:PEAK:TABLe:STATe?
Example
CALC:MARK:PEAK:TABL:STAT ON Turns on and displays the peak table.
Dependencies
When the Peak Table turns on, if Peak Threshold is On then it becomes the active function.
Preset
OFF
State Saved
Saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Peak Sort
Sets the peak table sorting routine to list the peaks in order of descending amplitude or ascending
frequency. The remote command can also be used to sort the peaks found using the
:CALCulate:DATA:PEAKs command.
Key Path
Peak Search, Peak Table
Remote Command
:CALCulate:MARKer:PEAK:SORT FREQuency|AMPLitude
:CALCulate:MARKer:PEAK:SORT?
Example
CALC:MARK:PEAK:SORT AMPL Sets sorting routine to list peaks in order of descending
amplitude.
CALC:MARK:PEAK:SORT?
Preset
AMPLitude
904
Remote Language Compatibility Measurement Application Reference