Amb (a minus b into a), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 232: Description, Syntax legacy products
5 Legacy Command Descriptions
AMB (A minus B into A)
AMB (A minus B into A)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Subtracts the points in Trace B from the corresponding points in Trace A, and sends the results
to Trace A. Thus, AMB can restore the original trace after an
"APB (Trace A Plus Trace B to A)" on
or a
"KSc (A Plus B to A)" on page 325
command has been executed.
The query AMB? returns different responses depending on the language being used. The 8560
Series languages return either a 1 or a 0 to indicate the On or Off status.The 8566, 8568 Series
languages all return either ON or OFF.
Format
AMB 0|1|OFF|ON
AMB?
Query Data Type
1 or 0, indicating ON or OFF state respectively.
SCPI Equivalent Commands
None
Preset
OFF
Couplings
Sets Trace B to View mode and turns
"AMBPL (A minus B plus Display Line into A)" on
(Normalize) OFF. All trace math is mutually exclusive, so turning one on turns the
other off and vice versa. Similarly, when AMB is on and you change Trace B to Clearwrite
or Maxhold, it turns AMB off.
Notes
The functions of the command AMB are identical to the command
232
Remote Language Compatibility Measurement Application Reference