Agilent Technologies Signal Analyzer N9030a User Manual
Page 600
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
I Skew
Sets the skew factor for the I channel. The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling.
Key Path
Input/Output, I/Q, I Setup
Remote Command
[:SENSe]:CORRection:IQ[:I]:SKEW
[:SENSe]:CORRection:IQ[:I]:SKEW?
Example
Delay the data for the I channel by 10 ns.
CORR:IQ:SKEW 10 ns
Preset
0
State Saved
Yes
This is unaffected by Preset but is set to the default value on a "Restore Input/Output Defaults"
or "Restore System Defaults->All"
Range
0 s to 100 ns
Min
0 s
Max
+100 ns
Initial S/W Revision
Prior to A.02.00
I Probe
Access the probe setup parameters for the I channel. See
Key Path
Input/Output, I/Q, I Setup
State Saved
No
Readback Text
[]
This is reporting the type of probe sensed on the I port. There is no parameter for overriding what
is sensed.
Initial S/W Revision
Prior to A.02.00
Attenuation
The attenuation is part of the calibration data stored with the probe type and is initially the
value that was returned by the last calibration. You can modify this value and any changes will
be stored with the calibration data and will survive power cycles and presets. When a probe
calibration is performed the attenuation value will be overwritten by the calibration.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe
600
Remote Language Compatibility Measurement Application Reference