6  RLC Swept SA Measurement Front-Panel & SCPI Reference

AMPTD Y Scale

Example

:POW:RANGe:OPT:ATT OFF

Initial S/W Revision

Prior to A.02.00

Elec Atten Only

Selects only the electric attenuator to participate in auto ranging. This offers less wear on the
mechanical attenuator and is usually faster.

This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.

Key Path

AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip

Example

:POW:RANGe:OPT:ATT ELEC

Initial S/W Revision

Prior to A.02.00

Mech + Elec Atten

In dual attenuator models, this selects both attenuators participate in the autoranging.

This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.

Key Path

AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip

Example

:POW:RANGe:OPT:ATT COMB

Initial S/W Revision

Prior to A.02.00

(Mech) Atten Step

This controls thestep size used when making adjustments to the input attenuation.

This key is labeled Mech Atten Step in dual attenuator models and Atten Step in single attenuator
models. In the dual attenuator configuration, this key only affects the step size of the mechanical
attenuator. 

Key Path

AMPTD Y Scale, Attenuation

Remote Command

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10 dB | 2 dB

[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?

Remote Language Compatibility Measurement Application Reference