6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
Example
:POW:RANGe:OPT:ATT OFF
Initial S/W Revision
Prior to A.02.00
Elec Atten Only
Selects only the electric attenuator to participate in auto ranging. This offers less wear on the
mechanical attenuator and is usually faster.
This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.
Key Path
AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip
Example
:POW:RANGe:OPT:ATT ELEC
Initial S/W Revision
Prior to A.02.00
Mech + Elec Atten
In dual attenuator models, this selects both attenuators participate in the autoranging.
This key is grayed out in measurements that do not support this functionality. The spectrum
analyzer measurement, Swept SA, does not support this functionality.
Key Path
AMPTD Y Scale, Attenuation, Pre-Adjust for Min Clip
Example
:POW:RANGe:OPT:ATT COMB
Initial S/W Revision
Prior to A.02.00
(Mech) Atten Step
This controls thestep size used when making adjustments to the input attenuation.
This key is labeled Mech Atten Step in dual attenuator models and Atten Step in single attenuator
models. In the dual attenuator configuration, this key only affects the step size of the mechanical
attenuator.
Key Path
AMPTD Y Scale, Attenuation
Remote Command
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement] 10 dB | 2 dB
[:SENSe]:POWer[:RF]:ATTenuation:STEP[:INCRement]?
Remote Language Compatibility Measurement Application Reference