6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe, Coupling
Readback Text
The last calibration date, or if no calibration exists, "(empty)".
Last:
Example:
Last: 8/22/2007
1:02:49 PM
Initial S/W Revision
Prior to A.02.00
Clear Calibration
Clears the calibration data for the current port and probe. It does not clear the data for other
probe types or other ports. If the sensed probe has EEPROM identification, only the data for that
specific probe is cleared. After this command has completed, the probe calibration state will be
the same as if no probe calibration had ever been performed for the specified channel and probe.
The probe attenuation will be the default value for that probe type and the Cable Calibration
frequency response corrections will be used. This command is dependent on the Differential Input
state. When Differential Input is on, both the data for the probe attached to the main port and the
data for the probe attached to the complementary port are cleared. When Differential Input is off,
only data for the probe attached to the main port is cleared.
Key Path
Input/Output, I/Q, I Setup | Q Setup, I Probe | Q Probe
Remote Command
:CALibration:IQ:PROBe:I|Q:CLEar
Example
Clear the calibration data for the I channel and the current probe (with EEPROM identification) or
probe type (without EEPROM identification).
:CAL:IQ:PROBe:I:CLE
Initial S/W Revision
Prior to A.02.00
Combined Differential/Input Z (Remote Command Only)
This is Remote Command only (no front panel) and is for backwards compatibility only. It
combines the Differential Input and Input Z selections into a single SCPI command.
Remote Command
:INPut:IMPedance:IQ U50|B50|U1M|B1M
:INPut:IMPedance:IQ?
Example
:INPut:IMPedance:IQ U50
This is equivalent to the following two SCPI commands:
602
Remote Language Compatibility Measurement Application Reference