Agilent Technologies Signal Analyzer N9030a User Manual
Page 1279
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TIME:RFPSelector:LCONducted?
Example
:CAL:TIME:RFPS:LCON?
State Saved
No
Restriction and Notes
Values are the date and time the last successful Align Now, 20 Hz – 30 MHz was executed. The
date is separated from the time by a semi-colon character.
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TEMPerature:RFPSelector:LCONducted?
Example
:CAL:TEMP:RFPS:LCON?
State Saved
No
Restriction and Notes
Value is in degrees Centigrade at which the last successful Align Now, 20 Hz – 30 MHz was
executed.
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TIME:RFPSelector:LRADiated?
Example
:CAL:TIME:RFPS:LRAD?
State Saved
No
Restriction and Notes
Value is the date and time the last successful Align Now, 30 MHz – 3.6 GHz was executed. The
date is separated from the time by a semi-colon character.
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode
All
Remote Command
:CALibration:TEMPerature:RFPSelector:LRADiated?
Example
:CAL:TEMP:RFPS:LRAD?
State Saved
No
Restriction and Notes
Value is in degrees Centigrade at which the last successful Align Now, 30 MHz – 3.6 GHz was
executed.
Remote Language Compatibility Measurement Application Reference
1279