6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
:CALCulate:MARKer[1]|2|3|4:FCOunt:RESolution:AUTO OFF|ON|0|1 is
accepted and ignored
:CALCulate:MARKer[1]|2|3|4:FCOunt:RESolution:AUTO?
Always returns 1
All of these commands cause the marker to become selected.
Preset
1Hz
ON
Initial S/W Revision
Prior to A.02.00
Couple Markers
When this function is On, moving any marker causes an equal X Axis movement of every other
marker which is not Fixed or Off. By “equal X Axis movement” we mean that we preserve the
difference between each marker’s X Axis value (in the fundamental x-axis units of the trace that
marker is on) and the X Axis value of the marker being moved (in the same fundamental x-axis
units).
Note that Fixed markers do not couple. They stay where they were while all the other markers
move. Of course, if a Fixed marker is being moved, all the non-fixed markers do move with it.
This may result in markers going off screen.
Key Path
Marker
Remote Command
:CALCulate:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:MARKer:COUPle[:STATe]?
Example
:CALC:MARK:COUP ON sets Couple Markers on.
Preset
Off, presets on Mode Preset and All Markers Off
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
All Markers Off
Turns off all markers. See Marker,
.
Key Path
Marker
Remote Command
:CALCulate:MARKer:AOFF
Example
CALC:MARK:AOFF turns off all markers.