6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
LXI
Opens a menu that allows you to access the various LXI configuration properties.
Key Path
System, I/O Config
Initial S/W Revision
Prior to A.02.00
LAN Reset
Resets the LAN connection.
Key Path
System, I/O Config, LXI
Initial S/W Revision
Prior to A.02.00
System IDN Response
This key allows you to specify a response to the *IDN? query, or to return the analyzer to the
Factory response if you have changed it.
To choose the factory-set response, press the Factory key.
To specify your own response, press the User key, and enter your desired response.
Key Path
System, I/O Config
Mode
All
Remote Command
:SYSTem:IDN
:SYSTem:IDN?
Notes
•
This affects the response given in all Modes of the Analyzer, unless the current Mode has
also specified a custom response, in which case the current Mode’s custom IDN response
takes precedence over the System’s, but only while that Mode is the current Mode..
•
It survives shutdown and restart of the software and therefore survives a power cycle
•
Null string as parameter restores the Factory setting
Preset
This is unaffected by Preset but is set to the original factory setting on a “Restore System
Defaults->Misc”
State Saved
No
Initial S/W Revision
A.06.00
1300