6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Changing the X Axis Unit will erase all your limit lines. Are you sure you want to do this? Press
Enter
or OK to proceed, or Cancel(Esc) to cancel.
Key Path
Meas Setup, Limits
Remote Command
:CALCulate:LLINe:CONTrol:DOMain FREQuency|TIME
:CALCulate:LLINe:CONTrol:DOMain?
Example
:CALC:LLIN:CONT:DOM FREQ deletes all currently existing limit lines, then sets all limit lines to
be specified in terms of frequency.
Couplings
This affects all limit lines simultaneously, and resets all limit line data except the .wav file and
email address stored in the Actions.
Preset
Freq, not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Delete All Limits
Deletes all limit lines. Pressing Delete All Limits purges the data from all limit line tables.
All limit data will be cleared and returned to factory preset settings.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK
key to delete all limits. Press ESC or Cancel to close this dialog.” The deletion is only performed
if you press OK or Enter; if so, after the deletion, the informational message “All Limits deleted”
appears in the MSG line.
Key Path
Meas Setup, Limits
Remote Command
:CALCulate:LLINe:ALL:DELete
Example
:CALC:LLIN:ALL:DEL deletes all data for all limit lines.
Initial S/W Revision
A.02.00
Limit Line Data (Remote Command Only)
Defines the limit line values, and destroys all existing data. Up to 200 points may be defined for
each limit using the following parameters.
Hz (for frequency) and seconds (for time).
Range: –30 Gs to +30 Gs for time limits, –3 kHz to +350 GHz for frequency limits.
Remote Language Compatibility Measurement Application Reference