Mksp (marker span), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 405: Description, Syntax legacy products

5 Legacy Command Descriptions
MKSP (Marker Span)
MKSP (Marker Span)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
This command operates only when the delta marker is On (see
"MKD (Marker Delta)" on page 385
or
"M3 [three] (Delta Marker)" on page 367
). When the delta marker is On and MKSP is executed,
the delta marker and active marker determine the start and stop frequencies. The left marker
specifies the start frequency, and the right marker specifies the stop frequency. If marker delta is
Off, there is no operation.
Format
MKSP
Query Data Type
N/A
SCPI Equivalent Commands
:CALCulate:MARKer2[:SET]:DELTA:SPAN (see
Notes
For 8566A/B, 8568A/B, The functions of MKSP are identical to
.
If the active marker is not a delta marker, there is no change in its position.
Remote Language Compatibility Measurement Application Reference
405