Agilent Technologies Signal Analyzer N9030a User Manual
Page 1424
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal1:SLOPe?
Example
TRIG:EXT1:SLOP NEG
Couplings
This same slope is used in the Ext1 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:SLOPe
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
External 2
Pressing this key, when it is not selected, selects the external input port that you will use for the
periodic frame trigger synchronization.
Pressing this key, when it is already selected, accesses the external 2 sync source setup menu.
Key Path
Trigger, Periodic Timer, Sync Source
Example
TRIG:FRAM:SYNC EXT2
Dependencies
In some models, there is no second External input. In these models, the External 2 key is
blanked and the EXTernal2 parameter will generate a “Hardware missing; Not available for this
model number” message.
Couplings
Same as External 2 trigger source.
Readback
External 2
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
1424
Remote Language Compatibility Measurement Application Reference