3 Programming the Analyzer
STATus Subsystem
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?
Example
STAT:QUES:CAL:EXT:NEED:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Needed Enable
This command determines which bits in the Questionable Calibration Extended Needed Condition
Register will set bits in the Questionable Calibration Extended Needed Event register, which also
sets bit 14 of the Questionable Calibration Register. The variable
decimal values of the bits you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Example
STAT:QUES:CAL:EXT:NEED:ENAB 2
Can be used to query if an EMI conducted alignment is needed.
Preset
32767
Min
0
Max
32767
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Needed Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Needed Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed[:EVENt]?
Example
STAT:QUES:CAL:EXT:NEED?