6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.
The relative trigger level is not available in some measurements. In those measurements the
RELative parameter, and the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
command (above), will generate an error if sent.
Dependencies
This key is grayed out and Absolute Trigger Level selected if the required hardware is not present
in your analyzer and the current measurement does not support Relative triggering.
Preset
–6 dB
GSM: –25 dB
State Saved
Saved in instrument state
Min
–45 dB
Max
0 dB
Default Unit
dB or dBc
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:RFBurst:LEVel
This legacy command is aliased to :TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the
PSA had ONLY relative burst triggering
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.04.00
Trigger Slope
It is set positive to trigger on a rising edge and negative to trigger on a falling edge.
Key Path
Trigger, RF Burst
Remote Command
:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:RFBurst:SLOPe?
Example
TRIG:RFB:SLOP NEG
Couplings
This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
1226
Remote Language Compatibility Measurement Application Reference