6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

need to send the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE command, above.

The relative trigger level is not available in some measurements.  In those measurements the
RELative parameter, and the :TRIGger[:SEQuence]:RFBurst:LEVel:TYPE
command (above), will generate an error if sent.

Dependencies

This key is grayed out and Absolute Trigger Level selected if the required hardware is not present
in your analyzer and the current measurement does not support Relative triggering. 

Preset

–6 dB

GSM:  –25 dB

State Saved

Saved in instrument state

Min

–45 dB

Max

0 dB

Default Unit

dB or dBc

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:RFBurst:LEVel

This legacy command is aliased to :TRIGger[:SEQuence]:RFBurst:LEVel:RELative because the
PSA had ONLY relative burst triggering

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.04.00

Trigger Slope

It is set positive to trigger on a rising edge and negative to trigger on a falling edge.

Key Path

Trigger, RF Burst

Remote Command

:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:RFBurst:SLOPe?

Example

TRIG:RFB:SLOP NEG

Couplings

This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).

Preset

POSitive

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe

Backwards Compatibility
Notes

The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.

Initial S/W Revision

Prior to A.02.00

1226

Remote Language Compatibility Measurement Application Reference