Agilent Technologies Signal Analyzer N9030a User Manual
Page 1334
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
the trace in Trace On (Update On and Display On), even if that detector was already selected.
Selecting a math mode other than Off for a trace (pressing the key or sending the equivalent
SCPI command) puts the trace in Trace On (Update On and Display On), even if that math mode
was already selected.
Loading a trace from a file puts that trace in View regardless of the state it was in when it was
saved; as does being the target of a Copy or a participant in an Exchange.
Initial S/W Revision
Prior to A.02.00
Trace Update State On/Off
Key Path
Trace/Detector
Remote Command
:TRACe[1]|2|...6:UPDate[:STATe] ON|OFF|0|1
:TRACe[1]|2|...6:UPDate[:STATe]?
Example
TRAC2:UPD 0 Makes trace 2 inactive (stops updating)
Couplings
Whenever you set Update to On for any trace, the Display is set to On for that trace.
Preset
1|0|0|0|0|0 (On for Trace 1; Off for 2–6)
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRACe:MODE VIEW
sets :TRACe:UPDate OFF, :TRACe:DISPlay ON, for the selected trace. In earlier analyzers, View
and Blank were trace modes, set by TRACe:MODE command. In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE VIEW command will yield its new equivalent, which is Update=Off, Display=On
Initial S/W Revision
Prior to A.02.00
Trace Display State On/Off
Key Path
Trace/Detector
Remote Command
:TRACe[1]|2|...6:DISPlay[:STATe] ON|OFF|0|1
:TRACe[1]|2|...6:DISPlay[:STATe]?
Example
TRAC2:DISP,1 Makes trace 2 visible
TRAC3:DISP,0 Blanks trace 3
Couplings
Whenever you set Update to On for any trace, the Display is set to On for that trace.
Preset
1|0|0|0|0|0 (On for Trace 1; Off for 2–6)
State Saved
Saved in instrument state
1334