6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See
.
Example
:CALC:MAM:DET2:DWEL 400 ms
Sets the dwell time for detector 2 to 400 ms
Preset
200 ms
State Saved
Saved in instrument state
Min
1 ms
Max
60 s
Default Unit
s
Initial S/W Revision
A.02.00
Detector 3 Dwell Time
This is the time specified by the user to dwell while taking the measurement for detector 3. The
minimum allowed dwell time is based on the current detector. If “Off” is selected for detector 3,
this key is grayed out and shows 200 ms.
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See
.
Example
:CALC:MAM:DET3:DWEL 400 ms
Sets the dwell time for detector 1 to 400 ms
Preset
200 ms
State Saved
Saved in instrument state
Min
1 ms
Max
60 s
Default Unit
s
Initial S/W Revision
A.02.00
BW & Avg Type
This key controls the type of bandwidth and average type coupling used in Measure at Marker.
If set to “Autocoupled”, then the RBW and Average Type are selected by the instrument during the
Measure at Marker function, according to the normal Autocouple rules, regardless of whether RBW
Remote Language Compatibility Measurement Application Reference