Agilent Technologies Signal Analyzer N9030a User Manual
Page 819
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Offset
Enters a menu which allows you to offset the limit trace by a specified frequency, time, or
amplitude. The offsets will be immediately applied to the limit trace for display and failure
calculation; the offset can also be applied to the points in the limit line.
Key Path
Meas Setup, Limits, Edit
Initial S/W Revision
A.02.00
X Offset
Offsets the limit trace by some specified frequency (for Frequency-based limit lines) or a time (for
time-based limit lines).
Key Path
Meas Setup, Limits, Edit, Offset
Remote Command
:CALCulate:LLINe[1]|2|...6:OFFSet:X
:CALCulate:LLINe[1]|2|...6:OFFSet:X?
Example
:CALC:LLIN:OFFS:X –50MHZ sets the X axis offset to –50 MHz.
:CALC:LLIN:OFFS:UPD will apply the X axis offset to all points in the limit line, then reset the X
axis offset to zero.
Preset
0 Hz if Limit X-Axis Unit is Frequency
0 S if Limit X-Axis Unit is Time
State Saved
Saved in instrument state, survives Preset
Min
–500 GHz
Max
500 GHz
Default Unit
Determined by X axis scale.
Initial S/W Revision
A.02.00
Y Offset
Offsets all segments in the limit line by some specified amplitude.
Key Path
Meas Setup, Limits, Edit, Offset
Remote Command
:CALCulate:LLINe[1]|2|...6:OFFSet:Y
:CALCulate:LLINe[1]|2|...6:OFFSet:Y?
Remote Language Compatibility Measurement Application Reference
819