6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

Offset

Enters a menu which allows you to offset the limit trace by a specified frequency, time, or
amplitude. The offsets will be immediately applied to the limit trace for display and failure
calculation; the offset can also be applied to the points in the limit line.

Key Path

Meas Setup, Limits, Edit

Initial S/W Revision

A.02.00

X Offset

Offsets the limit trace by some specified frequency (for Frequency-based limit lines) or a time (for
time-based limit lines).

Key Path

Meas Setup, Limits, Edit, Offset

Remote Command

:CALCulate:LLINe[1]|2|...6:OFFSet:X

:CALCulate:LLINe[1]|2|...6:OFFSet:X?

= if Limit X-Axis Unit is Frequency, =

Example

:CALC:LLIN:OFFS:X –50MHZ sets the X axis offset to –50 MHz.

:CALC:LLIN:OFFS:UPD will apply the X axis offset to all points in the limit line, then reset the X
axis offset to zero.

Preset

0 Hz if Limit X-Axis Unit is Frequency

0 S if Limit X-Axis Unit is Time

State Saved

Saved in instrument state, survives Preset

Min

–500 GHz

Max

500 GHz

Default Unit

Determined by X axis scale.

Initial S/W Revision

A.02.00

Y Offset

Offsets all segments in the limit line by some specified amplitude.

Key Path

Meas Setup, Limits, Edit, Offset

Remote Command

:CALCulate:LLINe[1]|2|...6:OFFSet:Y

:CALCulate:LLINe[1]|2|...6:OFFSet:Y?

Remote Language Compatibility Measurement Application Reference