3  Programming the Analyzer

STATus Subsystem

Preset

32767

Min

0

Max

32767

Status
Bits/OPC
dependencies

Sequential command

Initial S/W
Revision

Prior to A.02.00

Questionable Integrity Event Query

This query returns the decimal value of the sum of the bits in the Questionable Integrity Event
register.

The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.

Mode

All

Remote Command

:STATus:QUEStionable:INTegrity[:EVENt]?

Example

STAT:QUES:INT?

Preset

0

Status Bits/OPC dependencies

Sequential command

Initial S/W Revision

Prior to A.02.00

Questionable Integrity Negative Transition

This command determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register bit has
a negative transition (1 to 0)
The variable is the sum of the decimal values of the bits that you want to enable.

Mode

All

Remote
Command

:STATus:QUEStionable:INTegrity:NTRansition

:STATus:QUEStionable:INTegrity:NTRansition?

Example

STAT:QUES:INT:NTR 8

Measurement ‘regained calibration’ Summary will be reported to the Integrity Summary of the Status

Remote Language Compatibility Measurement Application Reference