3 Programming the Analyzer
STATus Subsystem
Preset
32767
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Integrity Event Query
This query returns the decimal value of the sum of the bits in the Questionable Integrity Event
register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity[:EVENt]?
Example
STAT:QUES:INT?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Negative Transition
This command determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register bit has
a negative transition (1 to 0)
The variable
Mode
All
Remote
Command
:STATus:QUEStionable:INTegrity:NTRansition
:STATus:QUEStionable:INTegrity:NTRansition?
Example
STAT:QUES:INT:NTR 8
Measurement ‘regained calibration’ Summary will be reported to the Integrity Summary of the Status
Remote Language Compatibility Measurement Application Reference