6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
Range (for SCPI command): 1 to 12. If the range is exceeded the value is clipped.
Couplings
The act of specifying the selected marker’s reference marker makes the selected marker a Delta
marker. If the reference marker is off it is turned on in Fixed mode at the delta marker location.
Preset
The preset default “Relative To” marker (reference marker) is the next higher numbered marker
(current marker +1). For example, if marker 2 is selected, then it’s default reference marker is
marker 3. The exception is marker 12, which has a default reference of marker 1.
Set to the defaults by using Restore Mode Defaults. This is not reset by Marker Off, All
Markers Off, or Preset.
State Saved
Saved in instrument state. Not affected by Marker Off and hence not affected by Preset or power
cycle.
Min
1
Max
12
Status Bits/OPC
dependencies
none
Default (selected when Restore Mode Defaults is pressed): next higher numbered marker or 1 if
marker 12.
Initial S/W Revision
Prior to A.02.00
Select Marker
Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.
Key Path
Marker
Notes
The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.
Preset
Marker 1
State Saved
The number of the selected marker is saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Select Marker
Specifies the selected marker. The term “selected marker” is used throughout this document to
specify which marker will be affected by the functions.
Key Path
Marker
Notes
The selected marker is remembered even when not in the Marker menu and is used if a Search
is done or a Band Function is turned on or for Signal Track or Continuous Peak.