Smooth (smooth trace), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 462: Description, Syntax legacy products
![background image](/manuals/40104/462/background.png)
5 Legacy Command Descriptions
SMOOTH (Smooth Trace)
SMOOTH (Smooth Trace)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Smooths the trace according to the number of points specified for the running average.
Each point value is replaced with the average of the values (in measurement units) of the given
number of points centered on it. Increasing the number of points increases smoothing at the cost
of decreasing resolution. If the number of points is an even number, then the number of points is
increased by one.
Smoothing decreases at the endpoints.
Format
SMOOTH TRA|TRB|TRC,
TRA corresponds to Trace 1, TRB corresponds to Trace 2, and TRC corresponds to Trace 3.
Query Data Type
N/A
SCPI Equivalent Commands
:TRACe:MATH:SMOoth TRACE(1|2|3|4|5|6) (see
)
:CALCulate:DATA:COMPress? ...
Notes
Prerequisite Commands:
when using trace data.
Some differences may be noticed between the smoothed trace in the legacy analyzers and
the smoothed trace using the same signal in X-Series instruments.
462
Remote Language Compatibility Measurement Application Reference