Agilent Technologies Signal Analyzer N9030a User Manual
Page 808
![background image](/manuals/40104/808/background.png)
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Relative to RL
Chooses whether the limit line amplitude points are coupled to the instrument reference level, and
whether the amplitude points are expressed as an offset from the instrument reference level.
For example, assume you have a limit line, and the reference level at –10 dBm. If Relative to RL is
“Off”, entering a limit line segment with an amplitude coordinate of –20 dB displays the limit line
segment at –20 dBm, and the limit line segment will not change amplitude if the reference level
amplitude changes. If Relative to RL is “On”, entering a limit line segment with an amplitude
coordinate of –20 dB displays the limit line segment at RL – 20 dB, or –30 dBm. Furthermore, if
the reference level amplitude changes to –30 dBm, the limit line segment will be displayed at RL –
20 dB, or –50 dBm.
It is possible to change this setting after a limit line has been entered. When changing from On to
Off or vice-versa, the amplitude values in the limit line table change so that the limit line remains
in the same position for the current reference level settings of the analyzer.
Key Path
Meas Setup, Limits, Properties, Fixed/Relative
Remote Command
:CALCulate:LLINe[1]|2|...6:AMPLitude:CMODe:RELative ON | OFF | 1 | 0
:CALCulate:LLINe[1]|2|...6:AMPLitude:CMODe:RELative?
Example
:CALC:LLIN:AMPL:CMOD:REL ON makes limit line 1 relative to the reference level amplitude.
Couplings
Pressing this button makes Reference level the active function.
Preset
Off, not affected by Mode Preset, preset by Restore Mode Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Description
Provides a description of up to 60 characters by which the operator can easily identify the limit.
Will be stored in the exported file. Can be displayed in the active function area by selecting as
the active function, if desired to be in a screen dump.
Key Path
Meas Setup, Limits, Properties
Remote Command
:CALCulate:LLINe[1]|2|...6:DESCription “Description”
:CALCulate:LLINe[1]|2|...6:DESCription?
808
Remote Language Compatibility Measurement Application Reference