6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
Example
:CALC:MAM:DET3 QPE
Sets the detector for measure at marker detector 1 to Quasi peak
:CALC:MAM:DET3 OFF
Sets the detector for measure at marker detector 1 to Off
Preset
EMI Average
State Saved
Saved in instrument state
Readback Text
Detector name
Backwards
Compatibility SCPI
[:SENSe]:EMI:MEASure:DETector:AVERage[:STATe] OFF|ON|0|1|
This command is included for compatibility with the E7400 and PSA option 239 . If sent with On
as a parameter, sets detector 3 to EMI Average. If sent with Off as a parameter, sets detector 3 to
Off
Initial S/W Revision
A.02.00
Detector 1 Dwell Time
This is the time specified by the user to dwell while taking the measurement for detector 1. The
minimum allowed dwell time is based on the current detector If “Off” is selected for detector 1,
this key is grayed out and shows 200 ms.
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See
Example
:CALC:MAM:DET:DWEL 400 ms
Sets the dwell time for detector 1 to 400 ms
Preset
200 ms
State Saved
Saved in instrument state
Min
1 ms
Max
60 s
Default Unit
s
Initial S/W Revision
A.02.00
Detector 2 Dwell Time
This is the time specified by the user to dwell while taking the measurement for detector 2. The
minimum allowed dwell time is based on the current detector. If “Off” is selected for detector 2,
this key is grayed out and shows 200 ms.