6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Autorange (Slower – Follows Signals)
Turns the ADC ranging to automatic which provides the best signal to noise ratio. Autorange is
usually preferred over the manual range choices.
Key Path
Meas setup, FFT IF Gain
Example
IF:GAIN:FFT AUTOrange
State Saved
Saved in instrument state.
Readback
Autorange
Initial S/W Revision
Prior to A.02.00
Low Gain (Best for Large Signals)
Forces FFT IF Gain to be off.
Key Path
Meas Setup, FFT IF Gain
Example
IF:GAIN:FFT LOW
State Saved
Saved in instrument state.
Readback
Low Gain
Initial S/W Revision
Prior to A.02.00
High Gain (Best Noise Level)
Forces FFT IF Gain to be on.
Key Path
Meas S0etup, FFT IF Gain
Example
IF:GAIN:FFT HIGH
Dependencies
The High setting for FFT IF Gain is grayed out when Swept IF Gain is manually set to Low (not
when Low is chosen by the auto-rules).
State Saved
Saved in instrument state.
Readback
High Gain
Initial S/W Revision
Prior to A.02.00