6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker To
Notes
Sending this command selects the subopcoded marker
Dependencies
This function is only available when the selected marker is a delta marker. Otherwise the key is
grayed out.
In addition, this function is not available when x-axis is the time domain
Couplings
All the usual couplings associated with setting Span apply (see “
”).
Backwards
Compatibility SCPI
:CALCulate:MARKer[1]|2|...12[:SET]:SPAN
Backwards Compatibility
Notes
In earlier ESA and PSA products, Mkr ->Span would adjust the span and change the delta
marker to a normal marker placing it at the center of screen. In all the X-Series products, this is
no longer true. The markers will remain in delta mode and the delta and reference marker will end
up on opposite edges of the display.
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference