6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Key Path
Trigger, TV
Remote Command
:TRIGger[:SEQuence]:TV:FMODe ENTire|ODD|EVEN
:TRIGger[:SEQuence]:TV:FMODe?
Example
TRIG:TV:FMOD EVEN
Notes
ODD is Field 1
EVEN is Field 2
Dependencies
This command is available only when Option B7B (TV trigger) is installed.
Preset
ENTire
Readback
Displays the Readback value
Initial S/W Revision
Prior to A.02.00
Entire Frame
When you select Entire Frame it causes the selected line number to be viewed as an offset into the
entire frame starting with line 1, the first line in Field One.
Key Path
Trigger, TV, Field
Example
TRIG:TV:FMOD ENT
Min
1, for all formats.
Max
525, for formats NTSC-M, NTSC-Japan, NTSC–4.43, PAL-M and PAL–60
625, for formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and SECAM-L.
Readback
Entire Frame
Initial S/W Revision
Prior to A.02.00
Field One
When you select Field One it causes the selected line number to be viewed as an offset into the
first field starting with Line 1, the first line in Field One.
Key Path
Trigger, TV, Field
Example
TRIG:TV:FMOD ODD
Min
Field 1 (ODD)
The minimum line is 1
Max
Field 1 (ODD)
Maximum line is 263 for formats NTSC-M, NTSC-Japan, NTSC–4.43, PAL-M and PAL–60
Remote Language Compatibility Measurement Application Reference