6  RLC Swept SA Measurement Front-Panel & SCPI Reference

System

Configure Applications - Remote Commands

The following topics provide details on using remote commands to configure the list of
applications you want to load into the instrument memory, or to query the Virtual Memory
utilization for your applications.

l

"Configuration list (Remote Command Only)" on page 1259

l

"Configuration Memory Available (Remote Command Only)" on page 1259

l

"Configuration Memory Total (Remote Command Only)" on page 1260

l

"Configuration Memory Used (Remote Command Only)" on page 1260

l

"Configuration Application Memory (Remote Command Only)" on page 1260

Configuration list (Remote Command Only)

This remote command is used to set or query the list of applications to be loaded in-memory.

Remote Command

:SYSTem:PON:APPLication:LLISt

:SYSTem:PON:APPLication:LLISt?

Example

:SYST:PON:APPL:LLIS "SA,BASIC,WCDMA"

Notes

Parameter uses the enum values of the
:INSTrument:SELect command.

The order of the is the order that the applications are
loaded into memory, and the order that they appear in the Mode Menu.

Error message –225 "Out of Memory" is reported when more applications are listed than can
reside in Virtual Memory. When this occurs, the existing applications load list is unchanged.

Preset

Not affected by Preset

State Saved

Not saved in instrument state

Initial S/W Revision

A.02.00

Configuration Memory Available (Remote Command Only)

Queries the amount of Virtual Memory remaining.

Remote Command

:SYSTem:PON:APPLication:VMEMory[:AVAilable]?

Example

:SYST:PON:APPL:VMEM?

Preset

Not affected by Preset

Initial S/W Revision

A.02.00

Remote Language Compatibility Measurement Application Reference

1259