6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
l
When Gate is on, Offset Res BW and Offset Video BW are ignored (if you set these values)
and the measurement works as if all Offset Res BW and all Offset Video BW are coupled
with the Res BW and the Video BW under the BW menu. When Gate is on, the Offset BW
key in the Offset/Limit menu is grayed out.
Preset
Off
State Saved
Saved in instrument state
Range
On|Off
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TIME:GATE[:STATe]
ESA compatibility
Backwards Compatibility
Notes
In ESA, Trig Delay (On) and Gate (On) could not be active at the same time. This dependency
does not exist in PSA or in X-Series.
Initial S/W Revision
Prior to A.02.00
Gate View On/Off
Turning on Gate View in the Swept SA measurement provides a single-window gate view display..
Turning on Gate View in other measurements shows the split-screen Gate View. In these
measurements, when the Gate View is on, the regular view of the current measurement traces and
results are reduced vertically to about 70% of the regular height. The Zero Span window, showing
the positions of the Gate, is shown between the Measurement Bar and the reduced measurement
window. By reducing the height of the measurement window, some of the annotation on the Data
Display may not fit and is not shown.
Key Path
Sweep/Control, Gate
Remote Command
[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0
[:SENSe]:SWEep:EGATe:VIEW?
Example
SWE:EGAT:VIEW ON turns on the gate view.
Dependencies
In the Swept SA measurement:
In Gate View, the regular Sweep Time key is grayed out . When pressed, the grayed out key puts
up the informational message "Use Gate View Sweep Time in the Gate menu."
In the other measurements:
When you turn Gate View on, the lower window takes on the current state of the instrument.
Upon leaving Gate View, the instrument takes on the state of the lower window.
When you turn Gate View on, the upper window Sweep Time is set to the gate view sweep time.
Couplings
These couplings apply to the Swept SA measurement:
Remote Language Compatibility Measurement Application Reference