6  RLC Swept SA Measurement Front-Panel & SCPI Reference
System

State Saved

No

Range

On | Off

Initial S/W Revision

Prior to A.02.00

Refresh

When pressed, refreshes the Show Errors display.

Key Path

System, Show, Errors

Initial S/W Revision

Prior to A.02.00

Clear Error Queue

This clears all errors in all error queues.

Note the following:

Clear Error Queue does not affect the current status conditions.

Mode Preset does not clear the error queue.

Restore System Defaults will clear all error queues.

*CLS only clears the queue if it is sent remotely and *RST does not affect any error queue.

Switching modes does not affect any error queues.

Key Path

System, Show, Errors

Initial S/W Revision

Prior to A.02.00

Input Overload Enable (Remote Command Only)

Input Overload errors are reported using the Input Overload status bit (bit 12 in the Measurement
Integrity status register).  Input Overloads (for example, ADC Overload errors) can come and go
with great frequency, generating many error events (for example, for signals just on the verge of
overload), and so are not put into the SCPI error queue by default.  Normally the status bit is the
only way for detecting these errors remotely.

It is possible to enable Input Overload reporting to the SCPI queue, by issuing the
:SYSTem:ERRor:OVERload ON command.   To return to the default state, issue the

1246