Standard event status enable – Agilent Technologies Signal Analyzer N9030a User Manual
Page 104
3 Programming the Analyzer
IEEE 488.2 Common Commands
Key Path
No equivalent key. Related key System, Show Errors, Clear Error Queue
Remote Command
*CLS
Example
*CLS Clears the error queue and the Status Byte Register.
Notes
For related commands, see the SYSTem:ERRor[:NEXT]? command. See also the STATus:PRESet
command and all commands in the STATus subsystem.
Status Bits/OPC
dependencies
Resets all bits in all event registers to 0, which resets all the status byte register bits to 0 also.
Backwards Compatibility
Notes
In general the status bits used in the X-Series status system will be backwards compatible with
ESA and PSA. However, note that all conditions will generate events that go into the event log,
and some will also generate status bits.
Initial S/W Revision
Prior to A.02.00
Standard Event Status Enable
Selects the desired bits from the standard event status enable register. This register monitors I/O
errors and synchronization conditions such as operation complete, request control, query error,
device dependent error, status execution error, command error, and power on. The selected bits
are OR’d to become a summary bit (bit 5) in the byte register which can be queried.
The query returns the state of the standard event status enable register.
Key Path
No equivalent key. Related key System, Show Errors, Clear Error Queue
Remote Command
*ESE
*ESE?
Example
*ESE 36 Enables the Standard Event Status Register to monitor query and command errors (bits 2
and 5).
*ESE? Returns a 36 indicating that the query and command status bits are enabled.
Notes
For related commands, see the STATus subsystem and SYSTem:ERRor[:NEXT]? commands.
Preset
255
State Saved
Not saved in state.
Min
0
Max
255
Status Bits/OPC
dependencies
Event Enable Register of the Standard Event Status Register.
Initial S/W Revision
Prior to A.02.00
104
Remote Language Compatibility Measurement Application Reference